Between Patient Maintenance for Midwest Plus
Maintenance Kit
Midwest Plus Aerosol Spray Maintenance Procedures for
XGT Between Patient Maintenance for Midwest Plus
Aerosol Spray XGT Only
Scrub
and Rinse
1. Disconnect handpiece from coupler.
2. Push
button to remove bur.
3. Scrub
and rinse handpiece under warm tap water with a soft sponge. While scrubbing and
rinsing, it is best to keep the handpiece in the upright position. Make sure the
fiber optic light port near the head of the handpiece is scrubbed well. Once a
week or as needed, gently rub alcohol-soaked cotton swab against the inside end
of the fiber optic element to remove debris or lubricant build
up.
Lubricate and Expel
1. While holding the
can of Midwest Plus Aerosol Spray upright, insert nozzle fully into
handpiece (without a bur).
2. Cover head of handpiece with Midwest spray guard or
paper towel. Lubricate your handpiece with one quick spray of Midwest
Plus then properly expel.
3. All lubricant must
be thoroughly expelled by running handpiece (without a bur) until it reaches
normal speed, or until all excess lubricant is expelled (a minimum of 20
seconds). Expel in operatory or using Midwest Air Station.
4. Wipe off any excess
lubricant on outside of handpiece with a dry gauze pad or paper towel. On a weekly basis
or if button gets stiff, spray Midwest Plus Aerosol into chuck for one
second.
|